What causes GC retention time shift?

What causes GC retention time shift?

A change in the temperature program often causes a retention time shift of all the peaks. A change in the initial temperature, the initial hold time, or the ramp rate can affect all of the peaks. Retention times increase with a lower initial temperature, longer initial hold time, or a slower ramp rate.

What is the working principle of GC?

The analysis performed by a gas chromatograph is called gas chromatography. Principle of gas chromatography: The sample solution injected into the instrument enters a gas stream which transports the sample into a separation tube known as the “column.” (Helium or nitrogen is used as the so-called carrier gas.)

What are the 4 principles of chromatography?

Four separation techniques based on molecular characteristics and interaction type use mechanisms of ion exchange, surface adsorption, partition, and size exclusion. Other chromatography techniques are based on the stationary bed, including column, thin layer, and paper chromatography.

Why does HPLC retention time shift?

Related to the last phenomenon are shifts in retention times that are caused by an increase of back-pressure in the column. Increasing back-pressure may indicate a contamination of the column, but even a clogged frit can affect retention times.

What is purge flow in GC?

The purge flow acts to keep the injector clean by flushing the upper portion with a low flow of gas. A small amount of the more volatile portion of the sample may escape from the injector via the purge vent. The higher the purge vent flow, the greater the potential loss of the more volatile sample compounds.

How do you increase peak shape in GC?

Steps that can be taken to improve early eluting peak shape:

  1. Use a split injection. This limits the amount of solvent that gets onto the column and reduces how much analyte dissolves in pooled solvent.
  2. Decrease the injection volume.
  3. Use a pressure pulsed injection.
  4. Use a guard column.
  5. Increase the column film thickness.

Why oxygen is not used in GC?

Whenever gases is used in the chromatography process, there’s a potential for gas leaks, whether from the supply lines, storage tanks, or from the chromatograph itself. Nitrogen gas displaces oxygen. If nitrogen were to leak, air levels would become deficient of oxygen and employees could suffer health problems.

What are the detectors used in gas chromatography?

7.1. Types of Detector

Detector Detectable Compound
General-Purpose Detectors
Flame ionization detector (FID) Organic compounds (other than formaldehyde and formic acid)
Thermal conductivity detector (TCD) All compounds other than the carrier gas
Barrier discharge ionization detector (BID) All compounds other than He and Ne

What is peak tailing in GC?

Peak tailing is a problem which is regularly encountered in capillary gas chromatography (GC). It can cause issues with resolution and peak integration, affecting both qualitative and quantitative analysis.

Why is my retention time shifting in GC?

GC Troubleshooting Series . Part Six: Retention Time Shifts Possible Cause: Changes in Column Dimensions vs. Previous Column. If your issue occurs after changing a column, it is probably due to changes in column length. The old column may have been trimmed over time to a significantly shorter length.

What does a chair-shaped GC peak indicate?

Figure 2 shows the ‘chair-shaped’ peak which is highly indicative of a poorly cut column, where the column entrance is partially blocked. Figure 2: Chair-shaped GC peaks which are indicative of problems with column inlet cut or partial blockage of the column, some examples of which are shown in the inset.

What is the cause of peak fronting in GC column?

In many cases the cause of peak fronting is an overload of the GC column produced by the injection of too great a mass of analyte.

Begin typing your search term above and press enter to search. Press ESC to cancel.

Back To Top